J Magn Reson. 1999 Jun;138(2):232-43 doi: 10.1006/jmre.1999.1727.

Pulsed electron-nuclear double resonance (ENDOR) at 140 GHz

Bennati M, Farrar CT, Bryant JA, Inati SJ, Weis V, Gerfen GJ, Riggs-Gelasco P, Stubbe J, Griffin RG.

Abstract

We describe a spectrometer for pulsed ENDOR at 140 GHz, which is based on microwave IMPATT diode amplifiers and a probe consisting of a TE011 cavity with a high-quality resonance circuit for variable radiofrequency irradiation. For pulsed EPR we obtain an absolute sensitivity of 3x10(9) spins/Gauss at 20 K. The performance of the spectrometer is demonstrated with pulsed ENDOR spectra of a standard bis-diphenylene-phenyl-allyl (BDPA) doped into polystyrene and of the tyrosyl radical from E. coli ribonucleotide reductase (RNR). The EPR spectrum of the RNR tyrosyl radical displays substantial g-anisotropy at 5 T and is used to demonstrate orientation-selective Davies-ENDOR.

PMID: 10341127