Am J Physiol Heart Circ Physiol. 2007 Jun;292(6):H3172-8 doi: 10.1152/ajpheart.01307.2006. 2007 Feb 23.

High-resolution imaging of murine myocardial infarction with delayed-enhancement cine micro-CT

Nahrendorf M, Badea C, Hedlund LW, Figueiredo JL, Sosnovik DE, Johnson GA, Weissleder R.

Abstract

The objective of this study was to determine the feasibility of delayed-enhancement micro-computed tomography (microCT) imaging to quantify myocardial infarct size in experimental mouse models. A total of 20 mice were imaged 5 or 35 days after surgical ligation of the left coronary artery or sham surgery (n=6 or 7 per group). We utilized a prototype microCT that covers a three-dimensional (3D) volume with an isotropic spatial resolution of 100 microm. A series of image acquisitions were started after a 200 microl bolus of a high-molecular-weight blood pool CT agent to outline the ventricles. CT imaging was continuously performed over 60 min, while an intravenous constant infusion with iopamidol 370 was started at a dosage of 1 ml/h. Thirty minutes after the initiation of this infusion, signal intensity in Hounsfield units was significantly higher in the infarct than in the remote, uninjured myocardium. Cardiac morphology and motion were visualized with excellent contrast and in fine detail. In vivo CT determination of infarct size at the midventricular level was in good agreement with ex vivo staining with triphenyltetrazolium chloride [5 days post-myocardial infarction (MI): r(2)=0.86, P

PMID: 17322414