Off-Resonance Imaging
Introduction:
The
development of superparmagnetic nanoparticles, such as monocrystalline
iron-oxide (MION), has been central to the development of magnetic resonance
molecular imaging. Iron-oxide based contrast agents have been used to both
label stem cells for cell tracking, as well as to detect specific and sparsely
expressed molecular targets. The strong magnetic dipole field generated by
these susceptibility agents both shifts the resonance frequency and decreases
the relaxation times of water molecules in the vicinity of the contrast agent.
These effects lead to a decrease in the MRI signal intensity in the vicinity of
the contrast agent on T2-weighted, T2*-weighted, and steady-state
free precession (SSFP) images.
The
detection of iron-oxides with sequences producing negative or dark signal
contrast has proven highly sensitive, capable of detecting contrast agent
concentrations in the nanomolar range. However, these techniques do suffer from
several potential problems: 1) extremely high local concentrations of MION may
produce a complete signal void precluding the accurate quantification of
nanoparticle concentration; 2) the detection of MION in regions with low
intrinsic signal-to-noise ratio (SNR) is difficult; 3) it is difficult to
distinguish MION induced susceptibility changes from those caused by
susceptibility artifacts such as air-tissue interfaces; 4) the long echo times
required to detect low concentrations of MION complicate the detection of
moving objects, such as the heart, where short echo times are optimal.
Several
off-resonance imaging (ORI) techniques have recently been proposed to create
positive contrast from iron-oxide agents and overcome some of the above
challenges. These techniques selectively image waters shifted off-resonance by
susceptibility agents using a variety of different methods including a gradient
dephasing technique [1,2], a spectrally selective RF technique [3,4], an
on-resonance water suppression method [5], and an off-resonance saturation
method [6].
These
positive contrast techniques have been used to successfully image high
concentrations of iron-oxide, such as those found in exogenously labeled stem
cells, at standard clinical field strengths [3,4]. However, there has been
little experience with these techniques in the imaging of more dilute
concentrations of iron-oxides or imaging at higher field strengths.
Furthermore, the quantitative relationships between the contrast agent concentration
and the induced chemical shifts, relaxation rates, and signal intensity have
not been fully elucidated.
The
aim of this study was thus to characterize in detail the performance of an
off-resonance positive contrast imaging technique for the detection of dilute
concentrations of iron-oxides, such as those produced by targeted molecular
probes, at high magnetic field strengths.
Materials and Methods:
1.
MION Phantoms: Three
different phantoms were made: (1) A MION phantom consisting of 3 mm NMR tubes
loaded with different concentrations of MION in aqueous solution; (2) An
air-water phantom consisting of eppendorf tubes filled approximately half way
with different concentrations of aqueous MION; (3) An air-fat phantom
consisting of an eppendorf tube filled with olive oil and water.
2.
In Vivo Mouse Model:
Myocardial macrophage infiltration was imaged in infarcted C57Bl6 mice (n = 4)
4 days after permanent ligation of the left coronary artery. This model has
been shown to produce a macrophage infiltrate in the healing infarct, beginning
48-72 hours after the initial injury. The mice were thus injected 48 hours
after coronary ligation with 15 mg/kg MION via the tail vein. Imaging was
performed a further 48 hours later to allow the probe sufficient time to be
taken up by the macrophages infiltrating the healing infarct. Uninfarcted mice
(n = 3) were used as control animals.
3.
Magnetic Resonance Imaging: The ORI pulse sequence used for both the in vitro and in vivo studies consisted of a variable
bandwidth on-resonance water excitation pre-pulse followed by gradient
crushers, a subsequent fat-selective excitation pre-pulse also followed by
gradient crushers, and a standard slice selective spin-echo imaging sequence.
MRI of the MION phantoms was performed on Bruker Avance 14 and 4.7 T scanners
using transmit/receive volume coils. T1 and T2 relaxation
times were determined for different aqueous MION concentrations from inversion
recovery spin-echo data and from multi-echo spin-echo data, respectively. In
vivo mouse ORI
studies were performed on a 4.7 T Bruker Avance scanner with a custom built
solenoid coil (5 turn coil, 30 mm diameter, 70 mm length). Default in vivo imaging parameters included: FOV = 35
x 30 mm, slice thickness = 1 mm, matrix = 128 x 128, on-resonance WSBW = 200
Hz, TR = 2000 ms, TE = 2.9 ms, and 4 signal averages.
4.
Quantification of Induced Chemical Shifts: The mean chemical shift induced by MION was quantified
by varying the on-resonance water suppression bandwidth (WSBW) and measuring
the mean signal intensity. The plot of the signal intensity versus WSBW was
then fit to a sigmoidal Boltzman function given by:
|
|
|||
|
|
|||
|
The
fit parameter x0 corresponds to the WSBW at which the signal intensity is half way
between the maximum (A1) and minimum (A2) values. This corresponds to twice the mean induced
chemical shift (d).
|
|
Results:
1. Aqueous MION Phantom:
Positive contrast images of an aqueous
MION phantom are shown in Figure 1. We note that the brightest MION solution
does not correspond to the highest MION concentration. Furthermore, a lower
MION concentration is detectable at the lower field strength.
|
|
|
Figure
1: Spin-echo
and off-resonance images of a MION phantom acquired at 4.7 T (top row, a-d)
and 14 T (bottom row, e-h). The phantom consists of five 3 mm NMR tubes, with
varying MION concentrations, immersed in an undoped water bath. The MION
concentrations range from 8-256 ug Fe/ml and are indicated in the respective
spin-echo images (a, e). The off-resonance images were acquired with water
suppression bandwidths (WSBW) of 250 (b, f), 500 (c, g), and 750 Hz (d, h).
The 4.7 T images were acquired with TE/TR = 5.6/1000 ms while the 14 T images
were acquired with TE = 3.5/1000 ms. |
2.
Quantification of MION Concentration:
Figure 2 clearly demonstrates that the ORI signal intensity
is non-linear. This is attributed to the very short T2 relaxation
times of the higher MION concentration samples. In contrast, the induced
chemical shift and the off-resonance transverse spin-relaxation rate (R2)
vary linearly with MION concentration.
|
|
ORI Signal Intensity |
Induced Chemical Shift |
Off-Resonance R2 |
|
4.7 T |
|
|
|
|
14 T |
|
|
|
|
Figure 2: Dependence of the
off-resonance signal intensity, induced chemical shift, and off-resonance R2 on
the MION concentration at 4.7 and 14 T. The mean ORI signal intensity was
determined for a water suppression bandwidth (WSBW) of 250 Hz. While the ORI
signal intensity is clearly non-linear, both the induced chemical shift and
the off-resonance R2 vary linearly with MION concentration. |
|||
3.
Susceptibility Shifts at Air-Water and Air-Fat Interfaces:
Significant susceptibility artifacts are observed at air
interfaces as shown in Figure 3. A broadband water suppression pulse is needed to
reduce or eliminate the off-resonance positive contrast at air interfaces.
|
|
|
Figure 3: Positive contrast
from susceptibility shifts at air-water (a-c) and air-fat (d-f) interfaces. The
images shown are: spin-echo image (a), off-resonance image with a 200 Hz WSBW
(b), off-resonance image with a 400 Hz WSBW (c), spin-echo image (d),
fat-suppressed spin-echo image (e), fat-suppressed off-resonance image with a
250 Hz WSBW (f). All images were acquired at 4.7 T with TE/TR = 5.6/4000 ms. |
4. In Vivo ORI of MION: Dependence
on the Water Suppression Bandwidth
Shown in Figure 4 are off-resonance images of normal and
infracted mice. The off-resonance images have been fused with traditional spin-echo
images. No off-resonance signal was observed in the myocardium of any of the
normal control mice. In contrast, clear evidence of positive contrast was
observed in the myocardium of all infarcted mice. Significant susceptibility
artifacts are evident in the thoracic wall. Increasing the WSBW reduces much of
the background off-resonance signal, however, the sensitivity to MION
accumulation in the myocardium is lost.
|
|
|
Figure 4: Off-resonance
imaging of normal (a, b) and infarcted (c, d) mice acquired at 4.7 T with
water suppression bandwidths (WSBWs) of 200 (a, c) and 400 Hz (b, d). The
off-resonance images have been mapped to a color scale and overlaid onto
conventional spin-echo images. Iron-oxide accumulation within macrophages
infiltrating the injured myocardium produces positive off-resonance contrast
(arrow) for a WSBW of 200 Hz, but not for 400 Hz. |
5. In Vivo ORI of MION: Dependence
on the Echo Time
Positive contrast (arrow) is
seen in the anterolateral myocardium at the shorter echo times but not at an
echo time of 8.7 ms. The use of short echo times is thus critical to preserve
the sensitivity of the off-resonance imaging technique.
|
|
|
Figure 5: ORI and spin-echo (SE) images of MION
accumulation within infarcted myocardium at 4.7 T. Fat suppressed spin-echo
images were acquired with echo times of 2.9 (a) and 11.6 ms (b). The negative
contrast enhancement produced by MION can be seen in the thinned
anterolateral wall (arrow). Off-resonance spin-echo images were acquired
using a water suppression bandwidth of 200 Hz and echo times of 2.9 (c) and
8.7 ms (d). |
6. ORI Contrast-to-Noise Ratio:
As shown in Figure 6, significantly greater CNR was seen
in the off-resonance images of the infarcted mice (n=4) than the control mice
(n=3), reflecting the known uptake of MION by macrophages infiltrating the
healing infarct.
|
|
|
Figure 6: Quantification of MION uptake by off-resonance
imaging. The contrast-to-noise ratio (CNR) between the infarcted
anterolateral myocardium and the uninjured septal wall is shown. |
Discussion:
The
MION induced chemical shifts are independent of the field strength. However, the
on-resonance water linewidth increases with increasing field strength (~15 Hz
at 4.7 T and ~45 Hz at 14 T) and provides a lower bound limit on the smallest
MION induced chemical shift that can be resolved. The optimal field for ORI
will thus be a compromise between maximal SNR and minimal linewidth. For our
experiments at 4.7 and 14 T, however, we have found that the spectral
linewidths were the limiting factor in detecting low MION concentrations. The
lowest MION concentration detectable in our aqueous MION phantom was 8 ug Fe/ml
at 4.7 T and 32 ug Fe/ml at 14 T. Increased sensitivity will likely be obtained
at lower field strengths, where linewidths are narrower and there is greater
spectral resolution between on- and off-resonance water molecules.
Due to
the short T2 relaxation times observed at high fields, the MION
concentration dependence of the ORI signal intensity is highly non-linear
(Figure 2). However, the off-resonance R2 and induced chemical shift
both vary linearly with MION concentration (Figure 2) and can be used for MION
quantification. Due to the short T2 relaxation times, the detection
of even relatively dilute MION concentrations depends critically on the
spin-echo time (Figure 4). Since ORI works optimally at very short echo times
it may be well suited for imaging MION in regions that experience motion, such
as the myocardium, where the longer TE’s required for T2*-weighted
imaging could result in significant motion artifacts.
Large
chemical shifts were consistently seen at air interfaces both in vitro (Figure 3) and in vivo (Figure 4). For example, the mean
induced chemical shift at the thoracic wall interface was 303 Hz, which is
approximately equivalent to the shift induced by an aqueous MION concentration
of 151 ug Fe/ml. Suppression of these non-specific background areas of positive
contrast could be achieved, but only through the use of broad water suppression
bandwidths. However, as shown in Figure 4, the use of broad water suppression
bandwidths suppresses signals from dilute concentrations of MION.
Conclusion:
Off-resonance
imaging of dilute concentrations of iron oxides can be successfully performed
at high field strengths but is complicated by the decreased spectral resolution
of on and off-resonance spins and the high R2 relaxation rates at
high fields. The optimal field strength for off-resonance imaging in small
animals, such as mice, will thus be a compromise between maximal SNR and
minimal linewidth.
References: