Quantitative Simultaneous PET-MR Imaging

Vertical Tabs

J Med Imaging (Bellingham)
2014 Jun 5
9083
908325
10.1117/12.2051578
Journal Articles
PubMed ID: 
25632403

Simultaneous PET-MR is a novel and promising imaging modality that is generating substantial interest in the medical imaging community, while offering many challenges and opportunities. In this study, we investigated whether MR surface coils need to be accounted for in PET attenuation correction. Furthermore, we integrated motion correction, attenuation correction, and point spread function modeling into a single PET reconstruction framework. We applied our reconstruction framework to in vivo animal and patient PET-MR studies. We have demonstrated that our approach greatly improved PET image quality.

Year: 
2014