Magn Reson Med. 2008 Jun;59(6):1355-64 doi: 10.1002/mrm.21585.

Fast slice-selective radio-frequency excitation pulses for mitigating B+1 inhomogeneity in the human brain at 7 Tesla

Zelinski AC, Wald LL, Setsompop K, Alagappan V, Gagoski BA, Goyal VK, Adalsteinsson E.

Abstract

A novel radio-frequency (RF) pulse design algorithm is presented that generates fast slice-selective excitation pulses that mitigate B+1 inhomogeneity present in the human brain at high field. The method is provided an estimate of the B+1 field in an axial slice of the brain and then optimizes the placement of sinc-like "spokes" in kz via an L1-norm penalty on candidate (kx, ky) locations; an RF pulse and gradients are then designed based on these weighted points. Mitigation pulses are designed and demonstrated at 7T in a head-shaped water phantom and the brain; in each case, the pulses mitigate a significantly nonuniform transmit profile and produce nearly uniform flip angles across the field of excitation (FOX). The main contribution of this work, the sparsity-enforced spoke placement and pulse design algorithm, is derived for conventional single-channel excitation systems and applied in the brain at 7T, but readily extends to lower field systems, nonbrain applications, and multichannel parallel excitation arrays.

PMID: 18506800