J Magn Reson Imaging. 2011 Sep;34(3):623-33 doi: 10.1002/jmri.22637. 2011 Jul 14.

Bone mineral imaged in vivo by 31P solid state MRI of human wrists

Wu Y, Reese TG, Cao H, Hrovat MI, Toddes SP, Lemdiasov RA, Ackerman JL.

Abstract

PURPOSE: To implement solid state (31)P MRI ((31)P SMRI) in a clinical scanner to visualize bone mineral.
MATERIALS AND METHODS: Wrists of seven healthy volunteers were scanned. A quadrature wrist (31)P transmit/receive coil provided strong B(1) and good signal-to-noise ratio (SNR). A (1)H-(31)P frequency converter was constructed to enable detection of the (31)P signal by means of the (1)H channel. Data points lost in the receiver dead time were recovered by a second acquisition with longer dwell time and lower gradient strength.
RESULTS: Three-dimensional (31)P images, showing only bone mineral of the wrist, were obtained with a clinical 3 Tesla (T) scanner. In the best overall case an image with isotropic resolution of ∼5.1 mm and SNR of 30 was obtained in 37 min. (31)P NMR properties (resonance line width 2 kHz and T(1) 17-19 s) of in vivo human bone mineral were measured.
CONCLUSION: In vivo (31)P SMRI visualization of human wrist bone mineral with a clinical MR scanner is feasible with suitable modifications to circumvent the scanners' limitations in reception of short-T(2) signals. Frequency conversion methodology is useful for implementing (31)P SMRI measurements on scanners which do not have multinuclear capability or for which the multinuclear receiver dead time is excessive.

PMID: 21761459